FOR PHOSPHOSENS® KIT CUSTOMERS

PhosphoSens® Kit Protocols and Resources

Your kit is here. Find the protocol, component list, plate reader settings and support for your kinase, phosphatase or cell lysate assay kit. Enter your catalog number below, or pick your kit.

How to read it: CSLS-AQT0101-KL-100 product line · target · format & size

Kinase Assay Kits

View ↓

Lysate Assay Kits

View ↓

Phosphatase Assay Kits

View ↓
Getting started

Before your first run

1 · Store

Store components right away

  • Move each tube to the storage temperature printed on its label
  • Keep the Europium solution for PhosphoSens-Red kits separate from kinetic reagents
  • Note the lot number on the box
2 · Check

Confirm your contents

  • Match tubes against the component list for your kit and size below
  • Something missing or damaged? Email orders@assayquant.com with your order and lot number
3 · Gather

Have these ready

  • Active kinase or phosphatase for recombinant kits. Shop AQT recombinant enzymes
  • Cell or tissue samples for lysate kits
  • Low-volume plate (Corning or ProxiPlate)
  • Plate seal: DOT Scientific T480
4 · Set up

Set up your plate reader

  • Kinetic kits: kinetic mode, Ex 360 / Em 492 nm, read every 30 s at 30 °C
  • Red kits: time-resolved endpoint, Ex 360 / Em 620 nm
  • Full settings and instrument guides are in Plate reader setup
CSKS · Recombinant kinase

Kinase Assay Kits

Continuous, real-time kinase activity with PhosphoSens-Kinetic, or endpoint TRF with PhosphoSens-Red. Kits ending in -KR are PhosphoSens-Red and include Europium.

Kits: CSKS-AQTxxxx-K-200 / -K-1000 (Kinetic), -KR-200 / -KR-1000 (Red) · Reagent Packs: CSKS-AQT-KRP-1 / -5, CSKS-AQT-RRP-1 / -5
Kinase assay products

What's in the box

Catalog # CSKS-AQTxxxx-K-200 (Kinetic) · CSKS-AQTxxxx-KR-200 (Red)Catalog # CSKS-AQTxxxx-K-1000 (Kinetic) · CSKS-AQTxxxx-KR-1000 (Red)Catalog # CSKS-AQT-KRP-1 (Kinetic) · CSKS-AQT-RRP-1 (Red). 10,000-assay packs: -KRP-5 / -RRP-5
ComponentVolume
SubstrateReagent Pack: pair with 1+ mg bulk substrate 60 µL 300 µL Not included
ATP Solution, 100 mM 60 µL 300 µL 600 µL
DTT Solution, 1 M 10 µL 50 µL 100 µL
Enzyme Reaction Buffer (ERB), 10X 600 µL 3,000 µL 6,000 µL
Enzyme Dilution Buffer (EDB), 5X Base 1,600 µL 8,000 µL 16,000 µL
EGTA Solution, 550 mM 10 µL 50 µL 100 µL
Europium, 30 mMRedPhosphoSens-Red kits (-KR) and Red Reagent Packs (-RRP) only 1,200 µL 6,000 µL 12,000 µL
CSLS · Cell & tissue lysates

Lysate Assay Kits

Measure endogenous kinase activity directly in lysates. Includes PhosphoPreserve extraction buffer and inhibitor cocktails. The Evaluation Kit adds controls and reference compounds for your first run. Protocols and validation reports for each target are on the Lysate Assays product page.

Kits: CSLS-AQTxxxx-KL-100 / -KL-1000 · Reagent Packs: CSLS-AQT-KRP-1 / -5
Lysate assay products

What's in the box

Catalog # CSLS-AQTxxxx-KL-100Catalog # CSLS-AQTxxxx-KL-1000Catalog # CSLS-AQT-KRP-1. 10,000-assay pack: CSLS-AQT-KRP-5
ComponentVolume
SubstrateReagent Pack: pair with 1+ mg bulk substrate 45 µL 450 µL Not included
ATP Solution, 100 mM 30 µL 300 µL 600 µL
DTT Solution, 1 M 5 µL 50 µL 100 µL
Enzyme Reaction Buffer (ERB), 10X 300 µL 3,000 µL 6,000 µL
Enzyme Dilution Buffer (EDB), 5X Base 800 µL 8,000 µL 16,000 µL
EGTA Solution, 550 mM 30 µL 50 µL 100 µL
PhosphoPreserve Cell Extraction Buffer, 1X Base 1,000 µL 3,000 µL 6,000 µL
PhosphoPreserve Phosphatase Inhibitor Cocktail-1, 100X 20 µL 60 µL 120 µL
PhosphoPreserve Phosphatase Inhibitor Cocktail-3, 60X 20 µL 60 µL 120 µL
PhosphoPreserve Protease Inhibitor Cocktail, 60X 20 µL 60 µL 120 µL
PhosphoControl PeptideEvaluation Kit only 10 µL Not included Not included
(+) Lysate ControlEvaluation Kit only 25 µL Not included Not included
(+) Recombinant Enzyme ControlEvaluation Kit only 2 µL Not included Not included
Reference Compound(s)Evaluation Kit only 5 µL Not included Not included
CSPS · Recombinant phosphatase

Phosphatase Assay Kits

Continuous phosphatase activity using phospho-Sox substrates. No ATP required. Kits ending in -KR are PhosphoSens-Red and include Europium.

Kits: CSPS-AQTxxxx-K-200 / -K-1000 (Kinetic), -KR-200 / -KR-1000 (Red) · Reagent Packs: CSPS-AQT-KRP-1 / -5, CSPS-AQT-RRP-1 / -5
Phosphatase assay products

What's in the box

Catalog # CSPS-AQTxxxx-K-200 (Kinetic) · CSPS-AQTxxxx-KR-200 (Red)Catalog # CSPS-AQTxxxx-K-1000 (Kinetic) · CSPS-AQTxxxx-KR-1000 (Red)Catalog # CSPS-AQT-KRP-1 (Kinetic) · CSPS-AQT-RRP-1 (Red). 10,000-assay packs: -KRP-5 / -RRP-5
ComponentVolume
SubstrateReagent Pack: pair with 1+ mg bulk substrate 60 µL 300 µL Not included
DTT Solution, 1 M 10 µL 50 µL 100 µL
Enzyme Reaction Buffer (ERB), 10X 600 µL 3,000 µL 6,000 µL
Enzyme Dilution Buffer (EDB), 5X Base 1,600 µL 8,000 µL 16,000 µL
EGTA Solution, 550 mM 10 µL 50 µL 100 µL
Europium, 30 mMRedPhosphoSens-Red kits (-KR) and Red Reagent Packs (-RRP) only 1,200 µL 6,000 µL 12,000 µL
PLATE READER SETUP

General Plate Reader Setup Guidelines

PhosphoSens delivers rich information using a simple add-and-read workflow and a standard fluorescence microplate reader. Below, we provide general instrument setup guidelines and downloadable Plate Reader Specific Guides. 

PhosphoSens-Kinetic General Plate Reader Setup

  • Use an instrument capable of reading fluorescence intensity in kinetic (continuous) mode at an excitation wavelength of ~360 nm (358-363 nm) and an emission wavelength of ~492 nm (485-498 nm). 

  • Readings can be made at your desired intervals and duration. We read every 30 seconds for 4 hours (or until assay plateaus).

  • All PhosphoSens Assays are developed and optimized to be run at 30°C.

  • Place your plate in the reader, and select a well to use for adjusting gain and focus. To adjust the gain, use a positive control (phosphopeptide) to avoid going off scale during the assay. When finished, select "Start Adjustment" (or equivalent) to calculate its optimal focal height and make any gain adjustments necessary.

PhosphoSens-Red General Plate Reader Setup

  • Use an instrument capable of reading an endpoint time-resolved fluorescence (RFU) reading with an excitation wavelength of ~360 nm and an emission wavelength  of ~620 nm. 

  • After the kinase reaction is complete, and Europium has been added, return your plate to the reader without it's plate seal, adjust the gain and read.

Plate Reader-Specific Setup Guides

For additional instrument-specific setup support, contact our support team.

INSTRUMENT SETUP GUIDE

BMG LABTECH PHERAstar/PHERAstarPlus /PHERAstar FS/PHERAstar FSX Microplate Readers Assay Setup Guide

DOWNLOAD
Next steps

Scale up, source enzymes or let us run it

Running more assays?

Reorder and scale up

Reagent Packs supply buffers and ATP for 2,000 or 10,000 assays. Pair them with bulk substrate for your target (1, 5, 10 or 20 mg). Order by the catalog number on your current kit.

Need active enzyme?

Recombinant kinases and phosphatases

Source the recombinant enzymes to pair with your PhosphoSens kit from the team that designed the assay.

Short on time or bandwidth?

Let our lab run it with KinSight™

Kinase and phosphatase selectivity profiling, compound testing, time-dependent inhibition and custom assay development, run on the same PhosphoSens platform.

FAQ

Kit FAQs

Answers to the questions we hear most in the first week with a new kit. More answers are in the full Support FAQ.

How should I store my kit components?

Store each tube at the temperature printed on its label as soon as the kit arrives. Lyophilized sensor peptide can be stored for up to 2 years at -20 °C or below. Once resuspended, aliquot it and store it at -20 °C for up to 1 year. Keep freeze-thaw cycles to a minimum.

My kit arrived warm or partly thawed. Can I still use it?

Store it right away, then email orders@assayquant.com with your order number, lot number and a photo of the package. We'll tell you whether it's usable.

Where do I find the Certificate of Analysis for my lot?

A printed Certificate of Analysis ships with every order. If you can't find it, email support@assayquant.com with your lot number and we'll send an electronic copy.

Which plates should I use?

We've validated three plates:

  • Corning 96-well Half Area White Flat Bottom NBS (#3642)
  • Corning Low Volume 384-well White Flat Bottom NBS (#3824)
  • ProxiPlate 384-shallow well Plus, White (#6008280)

Seal plates with DOT Scientific T480 for kinetic reads.

What reaction conditions and enzyme concentration should I start with?

Most PhosphoSens assays use standard conditions: 54 mM HEPES pH 7.5, 1 mM ATP (or the ATP Km), 1.2 mM DTT, 0.55 mM EGTA, 0.012% Brij-35, 10 mM MgCl₂, 10 µM sensor peptide and 0.05–10 nM kinase. Titrate your enzyme across that range first. Some targets need extra co-factors, so check your kit protocol or the target-specific conditions on the product support pages.

Should I run the assay at room temperature?

We run all PhosphoSens assays at 30 °C. Room temperature varies from lab to lab and season to season, and a cooler reaction runs slower, which can look like weak signal. Use a temperature-controlled reader if you can.

How much DMSO and other additives can the assay tolerate?

These concentrations cause less than 10% inhibition of the PhosphoSens signal. Stay at or below them:

  • DMSO 10%
  • CaCl₂ 2.5 mM
  • DTT 5 mM
  • EDTA 1 mM or EGTA 2 mM
  • MnCl₂ 250 µM
  • NaCl 150 mM
  • Na₃VO₄ 40 µM
  • β-glycerophosphate 10 mM
  • SDS 0.01%, Triton X-100 1%

Does my enzyme source matter?

Yes. Activity varies a lot between vendors and constructs (purity, tags, full-length vs. truncated). If your signal is weak, compare enzyme preparations side by side in the kinetic format. You can also source enzyme directly from AssayQuant's recombinant kinases and phosphatases.

When do I add Europium in a PhosphoSens-Red assay?

Add Europium after the kinase reaction is complete. Remove the plate seal, return the plate to the reader, adjust the gain and read in time-resolved mode (Ex ~360 nm / Em ~620 nm).

My signal-to-background is low. What should I check?

Start with three causes:

  1. The enzyme may be too dilute, unstable or not fully activated. Some kinases need co-factors, such as Ca²⁺ or lipid for PKCs, or Ca²⁺/calmodulin for some CAMKs.
  2. ATP and Mg²⁺ may need titrating for your peptide.
  3. Your samples may contain compounds that interfere with fluorescence. Run a background scan before the kinetic read.

A 1.5-fold increase in fluorescence is generally enough for a Z′ above 0.8.

Background fluorescence creeps up during a long read. Why?

This is usually evaporation. Seal the plate with an optically clear film that allows top reading. We recommend DOT Scientific T480.

Should I reorder a Discovery Kit or a Reagent Pack, and how much substrate do I need?

Discovery Kits include substrate for 1,000 assays. Reagent Packs supply buffers and ATP for 2,000 assays (-KRP-1 / -RRP-1) or 10,000 assays (-KRP-5 / -RRP-5) but no substrate, so pair them with bulk sensor peptide, available in 1, 5, 10 and 20 mg. At 10 µM in a 20 µL 384-well reaction, 1 mg of an average sensor peptide (about 2,150 g/mol) runs about 2,250 assays. To order, contact your Area Business Manager or orders@assayquant.com.

CONTACT

Talk to a scientist

Questions about setup, your data or a different plate reader? Send us a note. You can attach your plate layout or raw export and our team will review it with you. Prefer email? support@assayquant.com · +1 (508) 804-3155

Cookie Settings / Do Not Sell or Share My Personal Information